This English version is provided for convenience. In case of any discrepancy, the French version prevails.
Home › Products › Weighing stations

Negative-pressure weighing stations and dispensing booths

Weighing a powder is one of the most exposing operations in a laboratory. The product is light, it lifts at the slightest air movement, and it is precisely the operator leaning over the balance who stands in the path of the emission.

Our negative-pressure stations address this risk through extraction and absolute filtration that carry particles away from the face. Extraction may be horizontal — POWDERJET, PROTEC I and II — or vertical, with the IV GEL SPACE EX. The choice depends on the geometry of the station and the operator's movements.

One difficulty specific to weighing deserves mention: airflow disturbs the balance. This is why the PPS can be fitted with a granite slab, which decouples the balance from fan vibration and makes weighing to 0.01 mg possible under airflow.

For allergenic powders, antibiotics and hormones, the PS(X)/EQL weighing booth provides triple protection through combined recirculation and extraction, with a Bag-in Bag-out system.

How to choose

The question is not which station, but who needs protecting, and from what. The answer very nearly selects the model on its own.

What decidesWhat it changes
Who must be protectedA negative-pressure station protects the operator: extraction and absolute filtration carry particles away from the face. If the preparation must also remain sterile, you need a station that protects the product too — the PPS combines both functions, as a Class II safety cabinet and a weighing station.
The direction of extractionHorizontal on the POWDERJET and the PROTEC I and II, vertical on the IV GEL SPACE EX. The choice follows the geometry of the workstation and the operator's movement, not a matter of principle.
The weighing accuracy requiredAirflow disturbs the balance: that is the difficulty specific to this application. The PPS can be fitted with a granite slab, which decouples the balance from ventilation vibration and makes weighing to 1/100 mg under airflow possible.
How hazardous the powder isFor allergenic powders, antibiotics and hormones, the PS(X)/EQL weighing booth provides triple protection through combined recirculation and extraction, with a Bag in Bag out system: the filter is removed without exposing the technician.
The volume handledA benchtop station suits occasional sampling. As soon as the operator has to step into the zone with a drum or a floor-standing balance, a weighing booth is required, built to measure from an approved drawing.

Applicable standards

There is no single product standard for a weighing station: which references apply depends on what the station protects.

ReferenceWhat it covers
ISO 14644Classification of airborne particulate cleanliness. It is the reference for laminar flow weighing booths such as the PS(X)/EQL.
EN 12469The European standard for microbiological safety cabinets. It applies to the PPS, which is also a Class II cabinet, and not to negative-pressure stations that protect the operator alone.
GMP Annex 1Grades A, B, C and D for controlled-atmosphere areas, where weighing forms part of sterile pharmaceutical production.
IQ / OQ / PQ qualificationInstallation, operation, performance, followed by periodic requalification.

Frequently asked questions

Can you weigh accurately under an airflow?
Yes, provided the problem is addressed. Airflow disturbs the balance, and that is the purpose of the granite slab offered on the PPS: it decouples the balance from ventilation vibration. Without that precaution, fine weighing under airflow is not reliable.
What is the difference between a weighing station and a microbiological safety cabinet?
A negative-pressure weighing station protects the operator from what is being handled. A Class II safety cabinet additionally protects the product, through a filtered laminar downflow, and falls under EN 12469. The PPS combines both functions in one unit.
Is a Bag in Bag out system needed for weighing powders?
As soon as the powder retained by the filter cannot be decontaminated before removal: cytotoxics, CMR substances, hormones, some antibiotics. The filter is then removed in a sealed bag. Without that arrangement, maintenance becomes the dangerous moment.
Does a negative-pressure station protect the product being handled?
Not by itself: the air entering the station is room air. If the product must stay protected, you need a unit that supplies filtered air over the work zone — a Class II safety cabinet, or a laminar flow booth.
Are weighing booths built to measure?
The PS(X)/EQL booth is built to measure, from a drawing approved before production. The POWDERJET, PROTEC I and II and IV GEL SPACE EX benchtop stations come in standard sizes.

Need advice on this range?

Our sales engineers will define the configuration best suited to your application, and our teams handle installation, commissioning and IQ / OQ / PQ qualification across France and Europe.

Request a quotation →

← All product categories

×

Request a quotation

Reply within two working days from your area sales engineer.
Our teams are available Monday to Thursday 7:30 am – 6:00 pm, and Friday 7:30 am – 3:00 pm (CET).

Your details are used solely to handle your request. Under the GDPR you may access, correct or delete them by writing to info@adslaminaire.com.