This English version is provided for convenience. In case of any discrepancy, the French version prevails.
Home › Safety cabinets › OPTIMALE
OPTIMALE — ADS Laminaire Complies with EN 12469
Documentation

OPTIMALE — Class II microbiological safety cabinet

Class II microbiological safety cabinet with motorised sash, suitable for handling pathogens of risk groups 1, 2 and 3. Automatic ECM control and maintenance carried out entirely from the front.

Request a quotation →

Key features

Overview

The OPTIMALE is a Class II microbiological safety cabinet designed to protect the operator, the product and the immediate environment from airborne contamination. It is suitable for handling pathogens of risk groups 1, 2 and 3.

Its design prioritises accessibility: all maintenance operations are carried out from the front, allowing installation in a row or against a partition without reserving rear access. The front sash is motorised and ECM control maintains constant airflow despite filter clogging.

The OPTIMALE 9 and OPTIMALE 12 are available with French NF marking.

Installation, commissioning and IQ / OQ / PQ qualification are carried out by our own teams. Our service department covers France and Europe.

Dimensions and specifications

Dimensions in millimetres. Each value gives the internal (usable) / overall figure where applicable. Custom sizes on request.

SpecificationOPTIMALE 9 NFOPTIMALE 12 NFOPTIMALE 16OPTIMALE 18
Width
internal / overall
921 / 971 mm1 226 / 1 276 mm1 531 / 1 581 mm1 836 / 1 886 mm
Height
internal / overall
650 / 1 400 mm
Depth
internal / overall
616 / 786 mm
Weight190 kg230 kg300 kg350 kg
Puissance1.4 kW1.4 kW1.5 kW1.5 kW
FiltrationG4 pre-filter + HEPA absolute filter (H14 class, EN 1822) — ULPA optional
StandardEN 12469 — Biotechnology: performance criteria for microbiological safety cabinets

Source: ADS Laminaire general catalogue 2025-2026. Specifications may change without notice.

Related products

← Back to safety cabinets · All products

×

Request a quotation

Reply within two working days from your area sales engineer.
Our teams are available Monday to Thursday 7:30 am – 6:00 pm, and Friday 7:30 am – 3:00 pm (CET).

Your details are used solely to handle your request. Under the GDPR you may access, correct or delete them by writing to info@adslaminaire.com.